19
12.2
Calibration check of conductivity in buffer solutions
To check the calibration of the sensor in buffer solutions, the sensor is placed in a container of the buffer
solution. Ensure that the sensor tips are centered in the container away from the edges and the bottom
of the container. The conductivity values displayed can vary depending on the position of the
conductivity sensor in the container of buffer solution.
Perform the calibration check as follows:
Shut the isolation valves to the controller plumbing assembly.
Place the Conductivity sensor in the buffer solutions.
Verify calibrations in at least two buffer solutions.
Re-install the sensor into the plumbing. Ensure locking ring is locked.
Restore
flow
to
plumbing assembly
12.3
Conductivity vs. ppm
The unit measures the conductivity of the water. The ppm of the water may be measured instead of
conductivity. If ppm is measured, use the following chart for an approximation of the conductivity level
and calibrate to the conductivity level that is closest to the ppm level that is measured. Remember this is
just an approximation because the ions that make up the conductivity may be different than the particles
that make up the ppm reading.
Conductivity vs. PPM Table
µS/cm
ppm
µS/cm
ppm
µS/cm
ppm
2
1
120
68
900
560
4
2.1
140
80
950
600
6
3.2
160
91
1000
630
8
4.2
180
100
1575
970
10
5.2
200
115
1575
1300
12
6.4
220
127
2500
1700
14
7.4
240
139
3000
1575
16
8.5
260
150
3400
2400
18
9.6
280
164
4000
2750
20
11.0
300
176
4500
3150
25
13.5
350
210
5000
3500
30
16.0
400
240
5500
3900
35
19.0
450
270
6000
4300