16
relays manually, view the process screens, and calibrate the controller. He cannot change any other
settings. In the
Technician
mode the operator has full access to all of the menus.
A password is required to change from a tight security level to a less restrictive security level. Each level
has its own factory-preset password (2222 for Technician, 1111 for Operator). If the controller is in the
View Only
or
Operator
mode just press the appropriate password on the keypad to change to a less
restrictive security mode.
The passwords can be changed to personalized passwords from the
Technician Level Menu
.
NOTE: IF YOU USE PERSONALIZED PASSWORDS, MAKE SURE THEY ARE RECORDED IN A
SAFE AND SECURE PLACE.
10
Starting Up the Controller
Once the Installation is complete it is time to start up the controller.
Initiate sample flow to the controller by opening the sample line isolation valves. Check for leakage.
Power up the controller by either turning on the circuit breaker or plugging the power cord into a 120
VAC receptacle.
It is best to initialize the whole controller to remove any settings that may be in the memory before
programming the controller. Refer to section 18.4 of this manual to initialize the controller.
If applicable, install each node in software.
Set the clock by following section 19
Set the high and alarms by following section 15
Calibrate the sensor by following section 12
Configure the relays for operation by following section 13.1
Verify operation of the controller before leaving the area.
11
Operation of the Controller
11.1
Process Screen