IX. CARE AND CLEANING, continued
Disinfection Instructions
After cleaning, the next step is to disinfect. Always both clean
and disinfect.
1. Using a fresh wipe, wring the wipe to remove excess liquid and wipe all
areas thoroughly.
2.
Keep area wet for 2 minutes to ensure disinfectant remains in contact
for a sufficient time to kill all bloodborne pathogens.
3. Allow to air dry completely.
Note:
It is important that the analyzer be
thoroughly dry before using.
4. The optical glass (C in picture) should be carefully wiped clean with an
alcohol wipe and dried with gauze to remove any residue from the
disinfectant.
5. Inspect the glass and ensure it is clean when held at different angles. If
it is not, repeat Step 4.
Following cleaning and disinfection inspect the analyzer for the following
signs of deterioration. These include:
Scratching on optical glass
Etching on optical glass
Liquid under optical glass
o
liquid intrusion, or
o
condensation
Loss of adhesion on optical glass
Liquid under display lens
Loss of adhesion on display lens
Deterioration of painted surfaces (polymer crazing, cracking, swelling, softening,
peeling, etc.)
Any loose parts
Stop using the analyzer and contact Customer Service for a replacement
analyzer immediately if you notice any signs of deterioration.
If you have any questions, call PTS Customer Service.