Liquidew I.S. User Manual
Michell Instruments
53
APPENDIX B
Register Map:
Function
Register Name
Texmate ASCII Modbus RTU
MOISTURE
RESULT
254
40515/40516
DEWPOINT_C
CH1
253
40517/40518
TEMPERATURE_C
CH2
252
40519/40520
DEWPOINT_F
CH3
251
40521/40522
TEMPERATURE_F
CH4
250
40523/40524
LAST_VIEW
TOTAL1
16
40529/40530
INPUT_VALUE
TOTAL2
17
40531/40532
cur_table
USER_MEMORY_1
5121
45121
UpdateFlag
USER_MEMORY_3
5123
45123
MIN_DP
USER_MEMORY_4
5124
45124
MAX_DP
USER_MEMORY_5
5125
45125
MIN_TEMP
USER_MEMORY_6
5126
45126
MAX_TEMP
USER_MEMORY_7
5127
45127
COMPOUND_1
USER_MEMORY_10
5130
45130
RATIO_1
USER_MEMORY_11
5131
45131
COMPOUND_2
USER_MEMORY_12
5132
45132
RATIO_2
USER_MEMORY_13
5133
45133
OP_SOURCE
USER_MEMORY_15
5135
45135
FAULT_OUTPUT
SP5_RESET_VALUE
119
40643/40644
SETPOINT_1
Alarm Set Point 1
6
40535/40536
SETPOINT_2
Alarm Set Point 2
7
40537/40538
SETPOINT_3
Alarm Set Point 3
8
40539/40540
SETPOINT_4
Alarm Set Point 4
9
40541/40542
ASCII Communications
Start
Character Address
Read/
Write
Register
Address
Separator
Data
Value
Message Terminator
s
0 to 255
r
or
w
1 to 65535
,
-9999999
to 9999999
$
•
Reading the dew-point value from a meter that has an address of 2:
Send: s2r253$
•
Reading the LBMMSCF from a meter that has an address of 5:
Send: s5r244$