Liquidew I.S. User Manual
Michell Instruments
39
OPERATION
3.6.2 Output Range ZERO Value Settings
0. 0 0
Z E R O
a. Use the
and
keys to adjust the
ZERO
value.
b. Press the
Select
key to confirm. The display will then change to the next
parameter, the
SPAN
point of the output range.
NOTE: The output range ZERO value will be in ppm
W
, dew point or temperature,
depending on the specific source (moisture signal, dew-point signal or
temperature signal) selected in Section 3.6.1. Also, if it is in temperature or
dew point, it could be either C(ºC) or F(ºF) depending on the unit selection in
the Main Reading Page (refer to Section 3.2.2). The value will be automatically
converted based on the new units selected.
3.6.3 Output Range SPAN Value Settings
0. 0 0
S P A N
a. Use the
and
keys to adjust the
SPAN
value.
b. Press the
Select
key to confirm. The display will then return to the Main
Reading Page.
NOTE: The output range SPAN value will be in ppm
W
, dew point or temperature,
depending on the specific source (moisture signal, dew-point signal or
temperature signal) selected in Section 3.6.1. Also, if it is in temperature or
dew point, it could be either C(ºC) or F(ºF) depending on the unit selection in
the Main Reading Page (refer to Section 3.2.2). The value will be automatically
converted based on the new units selected.
3.7
Analog Output 2 (Sample Temperature - Only)
Analog Output 2 has a fixed temperature signal of -20 °C...+70 °C (-4 °F...+158 °F) and
cannot be configured.