INSTALLATION
Preparation of the Sensor Cable
In order to comply with hazardous area certification of the
product it is essential that the crimps/wire ferrules supplied
must be attached on to any cable installed into the connector.
a. As shown in
Figure 15 below, the crimps/wire ferrules should be applied
so that there is no possibility of a conductor strand of a core becoming
free.
When the crimp/wire ferrules are applied they should have a minimum of
two positions of crimping. After the crimp/wire ferrules are applied they
should be trimmed to a length of 5mm (0.2”) (see
Figure 16).
10
mm
Figure 15
Crimped Wires
Figure 16
Cut to 5mm
b. Cable connection to the dew-point transmitter is made via the terminal
block (4) (see
Figure 17). Remove the terminal housing lid (2) to access.
1
4
5
1
3
2
Figure 17
Dew-Point Transmitter Housing