Liquidew I.S. User Manual
24
97092 Issue 5.3, September 2021
INSTALLATION
2.4.3 Sensor Assembly Wiring
NOTE: If the analyzer has been ordered with a sampling system, the Liquidew
I.S. sensor assembly will be factory-wired to the junction box. In that case
disregard the following instructions and go to Section 3.
2.4.3.1 Dew-point Transmitter Wiring
HAZARDOUS AREA INFORMATION
The dew-point transmitter (Easidew PRO I.S.) is certified
intrinsically safe for use in hazardous areas.
Before using the Easidew PRO I.S. in any hazardous
environment ensure that personnel are completely familiar
with the above standards relating to the certification of this
instrument and also with the further information relating to
intrinsically safe apparatus to be found in standard EN60079-
14:1997 or equivalent, and up-to-date codes of practice in the
country of installation.
Installation of the Easidew PRO I.S. MUST be as per
system drawing in order to comply with the intrinsic safety
regulations.
Refer to ATEX/
QPS
/IECEx certificates for the dew-point and
temperature transmitters’ connection cable requirements
which stipulate maximum permissible mutual capacitance and
inductance to resistance ratio.