Liquidew I.S. User Manual
38
97092 Issue 5.3, September 2021
OPERATION
3.6
Analog Output 1 (Moisture Content or Dew Point - User-Selectable)
The source and scale of the 4...20 mA output can be configured in the
OUTPUT SETUP
menu.
From the Main Reading Page, press
Menu
to enter the main menu. Use the
key to
scroll to the
OUTPUT SETUP
option.
O U T P U T
S E T U P
3.6.1 Output Source Selection
Press
Select
to enter the
OUTPUT SETUP
sub menu. The first parameter is the mA
output signal source. The display will then change to the name of the currently selected
source. The full name of the output source will scroll across the bottom line of the
display.
M O I S T
S O U R C E
M O I S T
M O I S T U R E
The user can select moisture signal, dew-point signal, or temperature signal as the
output source.
a. Change the selection by scrolling through these three options using the
and
keys.
b. When the required source is shown on the display, press the
Select
key
to confirm. The 4...20 mA output will then switch to the selected source.
The display will then change to the next parameter, the ZERO point of the
output range.