Liquidew I.S. User Manual
42
97092 Issue 5.3, September 2021
OPERATION
3.9
Alarm Setting
There are 4 built-in alarm relays which can be configured independently. From the Main
Reading Page, press and hold the
Select
key then press the
key to enter the ‘ALARM
SETTING’ page.
0. 0 C
S P _ 1
3.9.1 Alarm Set-Point Adjustment
The first page, as shown above, in ‘ALARM SETTING’ is for adjusting the Set-Point of
Alarm Relay 1.
a. Use the
and
keys to adjust the set-point.
b. Press the
Select
key to confirm. The display will then show the Alarm
Set-Point for Alarm Relay 2.
c. Repeat steps (a) and (b) to adjust the set-point for Alarm Relay 2, Alarm
Relay 3 and Alarm Relay 4 accordingly.
d. After Alarm Set Point SP_4 for Alarm Relay 4 is set, press the
Select
key
to confirm. The display will then show the Alarm Control Setting Page
SPC_1 for Alarm Relay 1.
NOTE: The Alarm Set Point could be in ppm
W
, dew point (ºC/ºF) or temperature
(ºC/ºF) depending on the source selection for the specific alarm (this can be
set in SPC function). Details of selecting the alarm source are explained in
Section 3.9.3. It should not be affected by unit selection at the Main Reading
Page, or source selection of the 4...20 mA output under
OUTPUT SETUP
.