36
EDDYSENSOR Operating Manual
The fuses should be checked if a front panel LED is not lit (F2, F3 or F4) or
if the equipment does not function (F1 or F5).
Fuses
Fuse F4 for
–15V supply
Fuse F3 for
+15V supply
Fuse F2 for
5V supply
In carrying out any work on the equipment, please observe the ESD
instructions given on the separate sheet at the beginning of this handbook.
The fuses should be checked if a front panel LED is not lit (F2, F3 or F4) or
if the equipment does not function (F1 or F5).
To do so, switch off the power and then:
a)
Remove the front module by loosening the four retaining screws.
b)
Pull the module carefully out of the housing.
c)
Remove the blown fuse(s).
d)
Insert the new fuse(s).
e)
Slide the front module into its slot in the housing and seat it firmly.
f)
Tighten up the four retaining screws again.
Fuse F1 for 115/
230V supply
Fuse F5 for 115/
230V supply
Replacing the fuses
PCB of the front module
The fuses for the voltage supplies are located on the front module PM1040F
of the power unit.
Power Unit
Front
Back
PM1040F
EDDYSENSOR from above