20
EDDYSENSOR Operating Manual
Switch settings
The locations of the various switches on the printed circuit boards of the
Delay Unit are clearly shown in the following diagrams to help in setting up
the unit.
The delay is set up on the OUT1 or OUT2 switch as required. A delay of 0
- 9999 encoder/time pulses is available.
Delay
Duration
Switch to select between
encoder and time
Prescaler
Switches to set
the delay
Switches to
set the signal
duration
Switch to select defect
type (always set to WO
for this equipment)
Front module -
Upper PCB
The duration of the pulses ranging from 100ms to 10s can be selected
using the WIDTH switches on the right of the row together with the scaling
factor (1s/0.1s) given by position 1 of the Channel switch (see below).
Duration
0.1s
1-9
100 - 900ms
0
1s
1s
1-9
1 - 9s
0
10s
Switch setting
(WIDTH switch)
Scaling factor
(Channel switch Pos. 1)