9
100
µ
S/cm
Permanent display 2
only with control
(w = setpoint)
D1C2-cond.-011-GB
Permanent display 1
mea val.
fd. fwd.
reg. val.
70 %
59 %
100
µ
S/cm
1000
µ
S/cm
w:
Positive values of setting variable: Increase conductivity
Negative values of setting variable: Reduce conductivity
Restricted Operating Menu / Description
Conductivity calibration
Measurement parameters
D1C2-cond.-030-GB
Access to all setting menus can be blocked with an access code !
meas. parameter
setting ?
mr = meas. range
cc = cell constant
2000
µ
S/cm
cc :
mr :
1106
µ
S/cm 25,0
°
C
mr 2 : 2000
µ
S/cm
cc 2 : 8.54 /cm
1106
µ
S/cm 25,0
°
C
8.54/cm
The measured value can be calibrated by changing the cell constant to the actual conductivity value
(arrow keys).
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Cell constant cc
depending on
0.0001/cm
0.0060/cm
0.1499/cm
cc can be adjusted for
sensor type
0.001/cm
0.150/cm
1.499/cm
all mr over the
used
0.01/cm
1.50/cm
12.00/cm
complete area
Cell constant calibration (cc)
D1C2-cond.-040-GB
calibration cc
probe in buffer!
accept value?
1538
µ
S/cm 8.54
calibration cc ?
PAR1
calibration cc
cc: 8.54 /cm
cond: 100
µ
S/cm
calibration cc
cc: 8.50 /cm
please wait!
BA_DM_138_09_06_GB.p65
21.09.2006, 11:22 Uhr
9