17
Complete Operating Menu / Description
Installation factor setting
Enter the installation factor of the installed sensor here (arrow keys!). Determine the installation factor using
the data from the sensor documentation and the installation geometry.
D1C2-cond.-041-GB
install. factor
setting ?
install. factor
1.00
100
µ
S/cm
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Installation factor
1.00
0.01
0.01
9.99
Type of sensor setting
When changing the sensor, press Enter to load the actuation parameters permanently stored in
DULCOMETER
®
D1C for this sensor type. Metering and control stop and the measured value-specific error
messages are deleted.
D1C2-cond.-044-GB
type of sensor
setting ?
type of sensor :
ICT 1
sensor data
are changed!
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Type of sensor
ICT 1
ICT 1
The default
ICT 1-IMA
actuation parameters
ICT 2 / CLS50
are loaded
ICT 3 / CLS52
IMPORTANT
• The limit values, the setpoints and the standard signal outputs are set to the default
values.
• Check the settings in all menus!
BA_DM_138_09_06_GB.p65
21.09.2006, 11:22 Uhr
17