Restricted Operating Menu / Description

To determine the precise cell constant (cc) of the sensor, place the sensor in a calibration solution (sample 1)
with a known conductivity and select the second menu option. The DULCOMETER

®

  D1C displays the

conductivity it has calculated with the updated parameters. Press Enter when the value is constant.
Select the next menu option and enter the conductivity of the calibration solution (arrow keys). Press Enter.
The D1C displays the recalculated cell constant and saves the value.
The setting menu is inactive when “PAR2“ appears.
During the calibration, the metering is reduced to the set basic load and control stops.
The limit value monitor and the fault diagnosis system are reset. The standard signal of the “measured value“
output is frozen.

Parameter record

This setting menu can be used to temporarily enable additional parameters of the second parameter record
PAR2 to be edited (in addition to the parameters from the setting menu “measurement parameter setting?“).
I.e. the “limit setting?“ and “mA output 1 (2) setting?“ setting menus are enabled to allow editing of PAR2
(designation PS2) until you exit the series of setting menus from “parameter record“ to “general setting“.

It is possible to switch between the parameter records for operation using the “feed forward control“ contact
input, e.g. during a flushing process (contact open = PAR1, contact closed = PAR2). When switching between
the parameter records, the current parameter record in the setting menu “measurement parameter
setting?”,  “limit setting?“  and  “mA output 1 (2) setting“  is replaced by the other parameter record
(designation PS1 or PS2).

There is no access to the calibration menus in the case of PAR2 (designation PS2).

NOTE

• The designation PAR1 indicates that the parameter record 1 is actively measuring (contact

input “feed forward control“ e.g. open).
The designation PS1 indicates that the parameter record 1 can be edited.
Exception: The parameter record 1 cannot be edited while it is actively measuring
(designation PAR1). If your controller switches the D1C to parameter set 1 while it is being
edited, the D1C returns to the permanent display.

• PAR2, no control, no standard signal outputs.

Grenzwert2 oben

 5.00 ppm

Grenzwert1 unten

 1.00 ppm

Only with parameter record 2

D1C2-cond.-042-GB

parameter record

setting  ?

PAR1

parameter record

PS1

 Possible values

Initial value

Increment

Lower value

Upper value

Remarks

Parameter record

PS1

PS1
PS2

Possible values

Initial value

Increment

Lower value

Upper value

Remarks

Conductivity

Measured value

0.1 

µ

S/cm

µ

S/cm

200 

µ

S/cm

Measuring range 200 

µ

S/cm

Calibration

µ

S/cm

µ

S/cm

2000 

µ

S/cm

Measuring range 2000 

µ

S/cm

solution (LF)

0.01 mS/cm

µ

S/cm

20 mS/cm

Measuring range 20 mS/cm

0.1 mS/cm

µ

S/cm

200 mS/cm

Measuring range 200 mS/cm

1 mS/cm

µ

S/cm

2000 mS/cm

Measuring range 2000 

µ

S/cm

BA_DM_138_09_06_GB.p65

21.09.2006, 11:22 Uhr

10

Summary of Contents for DULCOMETER D1C Type D

Page 1: ...ductivity ProMinent Please completely read through the operating instructions Do not discard The warranty shall be invalidated by damage caused by operating errors Please enter the identity code of yo...

Page 2: ...Temperature via terminal Pt100 4 Manual temperature input Feed forward control 0 None 4 Flow as 0 4 20 mA standard signal and parameter record switching 5 Parameter record switching Control input 0 N...

Page 3: ...the series DULCOMETER D1C controller provide detailed safety information and are divided into clear steps IMPORTANT Please observe the parts of these operating instructions applicable to your particul...

Page 4: ...To increase a displayed numerical value and to change variables flashing display BRANCH BACK Button Back to permanent display or to start of relevant setting menu DOWN Button To decrease a displayed n...

Page 5: ...ation of pause 4 4 Feed Forward Control The D1C controller can process a signal of a feed forward control Depending on the device version see identity code and the setting this signal can be obtained...

Page 6: ...left Symbol Control on left Metering pump 2 Reduce conductivity Symbol Control off right Symbol Control on right Solenoid valve 1 Increase conductivity Symbol Control off left Symbol Control on left S...

Page 7: ...1 6 Operation diagram Access code correct Parameter setting Permanent display 1 Permanent display 2 Various Access code Setting menus The various menus are selected with the CHANGE button The menu is...

Page 8: ...c mr 1106 S cm 25 0 C 8 54 cm Number and scope of setting menus depends on the device general setting information alarm relais active meas parameter setting mr meas range cc cell constant Ti Td xp 0 s...

Page 9: ...cell constant 2000 S cm cc mr 1106 S cm 25 0 C mr 2 2000 S cm cc 2 8 54 cm 1106 S cm 25 0 C 8 54 cm The measured value can be calibrated by changing the cell constant to the actual conductivity value...

Page 10: ...rol contact input e g during a flushing process contact open PAR1 contact closed PAR2 When switching between the parameter records the current parameter record in the setting menu measurement paramete...

Page 11: ...n 5 00 ppm Grenzwert1 unten 1 00 ppm limit PS1 setting limit2 upper 1500 S cm limit1 lower 1000 S cm Limit values Restricted Operating Menu Description Control D1C2 cond 015 GB Ti Td xp 0 s 10 xp 10 3...

Page 12: ...0 mS cm 210 mS cm Measuring range 200 mS cm 1000 mS cm 1 mS cm 100 mS cm 2100 mS cm Measuring range 2000 mS cm Control with dead zone 2 setpoints necessary Setpoint 2 Setpoint 1 Adjustment of measurin...

Page 13: ...ing correction value setting type of sensor setting install factor setting cali zero point PAR1 calibration cc PAR1 calibration parameter record setting PAR1 general setting information Only with serv...

Page 14: ...nductivity value arrow keys The precondition is a known temperature coefficient and constant temperature Metering and control stops if the measurement range is changed Setpoints limit values and stand...

Page 15: ...onstant cc of the sensor place the sensor in a calibration solution sample 1 with a known conductivity and select the second menu option The DULCOMETER D1C displays the conductivity which it has calcu...

Page 16: ...ed value or correction value output is reduced to 0 4 mA IMPORTANT Only values determined at 25 C may be entered as conductivity values The calibration must be carried out with the same solution at bo...

Page 17: ...e of sensor setting When changing the sensor press Enter to load the actuation parameters permanently stored in DULCOMETER D1C for this sensor type Metering and control stop and the measured value spe...

Page 18: ...n value calibration temp 33 0 C only with correction variable Correction value Complete Operating Menu Description Possible values Initial avalue Increment Lower value Upper value Remarks Type of temp...

Page 19: ...wer control Possible values Initial value Increment Lower value Upper value Remarks Relay adjustment as per Motor At limit value the identity code Solenoid valve SV1 SV2 relays remain active Limit val...

Page 20: ...mallest permitted operating factor of the connected device The actuating variable determines the ratio ton cycle and thus the switching times see fig above The min time influences the switching times...

Page 21: ...ing menus meas parameter setting limit setting and mA output 1 2 setting is replaced by the other parameter record designation PS1 or PS2 There is no access to the calibration menus in the case of PAS...

Page 22: ...ime limits t on off 1 s 1 s off 9999 s Results in message and alarm off 0 s Function switched off no message no alarm Switch on delay 0 s 1 s 0 s 9999 s Only available in t on complete operating menu...

Page 23: ...he control range places a maximum limit on the control range used IMPORTANT The servomotor must be actuated with the same care as for the calibration of a measuring sensor To ensure correct function t...

Page 24: ...crement Lower value Upper value Remarks Control normal normal When controlling with dead with dead zone zone the regulated value is manual not used for measured values within the dead zone Setpoint se...

Page 25: ...trol regulated value mA output 1 regulated value measured value mA output 1 measured value mA output 1 100 0 C 0 mA 20 mA correction value 0 0 C 0 0 mS cm 200 0 mS cm Possible values Initial value Inc...

Page 26: ...relay active active closed parameter set 2 Access to all setting menus can be blocked with an access code Possible values Initial value Increment Lower value Upper value Remarks Variable allocation a...

Page 27: ...alue or correction value are not affected by the pause After pause is activated the operating outputs remain at 0 for the length of the time delay td The time delay td must be set up in such a way tha...

Page 28: ...ded drops below value Check feed forward yes Signal 3 0 0 2 mA Check sensor transformer input or 23 0 2 mA and cable connection Signal below value multiplicative Stop yes The last valid value is used...

Reviews: