22
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Type of limit
Limit 1
lower
upper
Limit transgression when
transgression
Limit 2
upper
lower
exceeding or dropping
off*
below value.
*only with limit value relay
Limit value1; Limit value 2
0; 2000 mS/cm
1 mS/cm
-100 mS/cm
2100 mS/cm
Meas. range 2000 mS/cm
0; 200 mS/cm
0.1 mS/cm
-10 mS/cm
210 mS/cm
Meas. range 200 ms/cm
0; 20 mS/cm
0.01 mS/cm
-1 mS/cm
21 mS/cm
Meas. range 20 ms/cm
0; 2000
µ
S/cm
1
µ
S/cm
-100
µ
S/cm
2100
µ
S/cm
Meas. range 2000
µ
s/cm
0; 200
µ
S/cm
0.1
µ
S/cm
-10
µ
S/cm
210
µ
S/cm
Meas. range 200
µ
s/cm
Hysteresis limit values
10 mS/cm
1 mS/cm
0 mS/cm
2100 mS/cm
Effective in direction of
1 mS/cm
0.1 mS/cm
0 mS/cm
210 mS/cm
”Cancelling limit trans-
0.1 mS/cm
0.01 mS/cm
0 mS/cm
21 mS/cm
gression”
10
µ
S/cm
1
µ
S/cm
0
µ
S/cm
2100
µ
S/cm
1.0
µ
S/cm
0.1
µ
S/cm
0
µ
S/cm
210
µ
S/cm
Checkout time limits t on
off
1 s
1 s/off
9999 s
Results in message and
alarm.
off = 0 s:
Function switched off,
no message, no alarm
Switch-on delay
0 s
1 s
0 s
9999 s
Only available in
∆
t on
complete operating menu
Switch-off delay
0 s
1 s
0 s
9999 s
Only available in
∆
t off
complete operating menu
Max. switch-on time
off
1 s
0 s/off
9999 s
Only available in
Limit value 1;
complete operating menu
Limit value 2
Function switched off
Complete Operating Menu / Description
Limit values
D1C2-cond.-024-GB
Access to all setting menus can be blocked with an access code !
Only with limit value relay
limit 1
lower
limit 1
lower
∆
t on
∆
t off
0 s
0 s
limit 2
upper
limit 2
∆
t on
∆
t off
0 s
0 s
hysteresis
limits:
10
µ
S/cm
checkout time
limits:
limit 1
time:
0 s
limit 2
time:
limits PS1
setting ?
1000
µ
S/cm
1500
µ
S/cm
1500
µ
S/cm
1000
µ
S/cm
max.
max.
0 s
off
upper
If the limit transgression is longer than the “delay time limit value“, an error message is triggered (requires
acknowledgement) and the alarm relay opens; if “control“ is also set to “off“, the control process will stop.