Error message/Warning
Condition
Remarks
Temperature range restricted
For the chosen temperature
xx – 100
°
C
coefficient
α
, a correct reading
can only be obtained for the
displayed temperature range
Temperature interval incorrect
∆
temperature
≥
10.0
°
C
∆
temperature
≤
50.0
°
C
Complete Operating Menu / Description
Calibration of temperature coefficient
α
The temperature coefficient
α
is determined again by means of a two point calibration. During the
calibration, the metering is reduced to the set basic load and control stops. The limit value monitor and the
fault diagnosis system are reset. The standard signal of the measured value or correction value output is
reduced to 0/4 mA.
IMPORTANT
Only values determined at 25
°
C may be entered as conductivity values. The calibration
must be carried out with the same solution at both temperatures.
D1C2-cond.-032-GB
Access to all setting menus can be blocked with an access code !
calibration at
probe in buffer !
temperature 1
1538
23,8
°
C
calibration at
probe in buffer !
temperature 2
1500
48,5
°
C
permanent display 1
calibration
α
cc :
α
:
8,54/cm
1,90 %/
°
C
calibration
α
cc :
α
:
8,50/cm
2,52 %/
°
C
calibration at
accept value ?
temperature 1
1538
23,8
°
C
calibration at
accept value ?
temperature 2
1500
48,5
°
C
calibration
temp1
value:
temp2
48,5
°
C
23,8
°
C
1500
calibration
temp1
value:
1500
23,8
°
C
PAR1
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Set temperature
Correction value
0.1
°
C
0
°
C
100
°
C
BA_DM_138_09_06_GB.p65
21.09.2006, 11:22 Uhr