TE-input
Type
Min. value
Max. value
Standard
B
E
J
K
L
N
R
S
T
U
W3
W5
LR
0°C
-100°C
-100°C
-180°C
-200°C
-180°C
-50°C
-50°C
-200°C
-200°C
0°C
0°C
-200°C
+1820°C
+1000°C
+1200°C
+1372°C
+900°C
+1300°C
+1760°C
+1760°C
+400°C
+600°C
+2300°C
+2300°C
+800°C
IEC 60584-1
IEC 60584-1
IEC 60584-1
IEC 60584-1
DIN 43710
IEC 60584-1
IEC 60584-1
IEC 60584-1
IEC 60584-1
DIN 43710
ASTM E988-90
ASTM E988-90
GOST 3044-84
Cold Junction compensation (CJC):
via external sensor in connector CJC. . . . . . . . . . . . . . .
20...28°C ≤ ± 1°C
-
20...20°C / 28...70°C ≤±2°C
via internal CJC-sensor
. . . . . . . . . . . . . . . . . . . . . . . . . ±(2,0°C + 0,4°C * Δt)
Δt = internal temperature-ambient temperature
Sensor error detection, all TE-types . . . . . . . . . . . . . . . Yes
Sensor error current:
When detecting
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Nom. 2 μA
else
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 μA
Current input
Measurement range. . . . . . . . . . . . . . . . . . . . . . . . . . . .0...23 mA
Programmable measurement ranges. . . . . . . . . . . . . . .0...20 and 4...20 mA
Input resistance
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Nom. 20 Ω + PTC 50 Ω
Sensor error detection:
Loop break 4...20 mA. . . . . . . . . . . . . . . . . . . . . . . . . . .Yes
Voltage input
Measurement range. . . . . . . . . . . . . . . . . . . . . . . . . . . .0..12 VDC
Programmable measurement ranges. . . . . . . . . . . . . . 0...1 / 0,2...1 / 0...5 /
1...5
/ 0...10 and 2...10 VDC
Input resistance
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Nom. 10 MΩ