VI. Troubleshooting (continued)
Symptoms
Causes and Comments
The machine runs but does not
advance to the “Run” screen.
A sensor for one of the heads may be failing, or a belt may be slipping. Please contact Promega
or your local Promega representative for service.
Poor sample quality, low yield
or low purity
See the appropriate Technical Bulletin or Technical Manual supplied with your Maxwell
®
16
Purification Kit for more information.
Expected methods are not
shown on the LCD screen
Verify the firmware setup. Refer to Section III.D.
Calibration error:
Error Code 1:
Error Code 1 indicates a Platform calibration error. (A picture of the instrument platform is shown
in Section II, Figure C, page 9.)
• Verify that there are no obstructions in the slide channels, in the rear, or in front of the platform
that prevent it from moving back and foward freely.
• If in LEV mode, ensure the cartridges are seated completely in the LEV Cartridge Rack, and
that the rack is seated onto the platform properly. Refer to Figure 4B (page 21) for a diagram.
• Turn the instrument off, wait a few seconds, then turn the instrument back on. If the calibration
error persists, please contact Promega or your local Promega representative for service.
Calibration error:
Error Code 2:
Error code 2 indicates a Plunger Bar calibration error (A picture of the plunger bar is shown in
Section II, Figures A and B, page 9.)
• Verify that there are no solid particulates (e.g., bone, swab tips, large particulates, etc.) inside
Well #1 of the cartridge. Solid particulates which are not easily macerated, may obstruct the
plunger from moving freely to the bottom of Well #1 during processing.
• Verify that the plastic reagent cartridge is seated properly. Refer to Figure 4, Panels A and B
(page 21) for a diagram.
If in LEV mode
, ensure the cartridges are seated completely in the
LEV Cartridge Rack, and that the rack is properly seated on the platform.
If in SEV mode
,
ensure the cartridges are seated properly onto the platform.
• If the error occurs during plunger loading, ensure the hardware configuration matches the
firmware settings. Select “No” after the “Continue?” prompt on the LCD screen (see Section
IV.B, page 20).
• Verify that the shipping anchors have been adjusted properly. Refer to Section III.B.
• Turn the instrument off, wait a few seconds, then turn the instrument back on. If the
calibration error persists, please contact Promega or your local Promega representative for
service.