(i)
Changing from Standard to LEV Configuration (continued)
Promega Corporation ·
2800 Woods Hollow Road
·
Madison, WI 53711-5399 USA
·
Toll Free in USA 800-356-9526
·
Phone 608-274-4330
·
Fax 608-277-2516
· www.promega.com
Part# TM274
Printed in USA.
Page 14
Revised 6/08
4.
Gently lift up the SEV Magnetic Rod Assembly and
remove it.
5.
Insert the black LEV Plunger Bar Adaptor onto the
bottom of the remaining arm. Make sure that the side
containing the set screws faces upward. Several
magnets will “click” and hold the LEV Plunger Bar
Adaptor in place once it is attached. The LEV Plunger
bar should then be fairly difficult to remove.
6.
Insert the black LEV Magnetic Rod Assembly. The set
screws should be facing toward you. Press firmly to
place the LEV Magnetic Rod Assembly in position. It
will fit tightly into place.
6391T
A
Orientation of the Magnetic Rod Assembly
front
back
7.
Secure the LEV Magnetic Rod Assembly with the three
thumbscrews from the standard magnetic rod
assembly, and hand-tighten. Proceed to Section III.D,
and complete the configuration by updating the
Firmware to the LEV (Low Elution Volume) settings.
Failure to change the instrument firmware
operational mode from the standard elution volume
to the low elution volume settings will cause damage
to the instrument.
SEV Magnetic Rod Assembly
LEV Plunger Bar Adaptor
!
tm274.0608.qxp 6/4/2008 9:51 AM Page 14