III.D. Operational Mode Setup (continued)
(i) Setting the Maxwell
®
16 Instrument Operational Mode
(ii) Changing the Maxwell
®
16 Instrument Operational Mode
Promega Corporation ·
2800 Woods Hollow Road
·
Madison, WI 53711-5399 USA
·
Toll Free in USA 800-356-9526
·
Phone 608-274-4330
·
Fax 608-277-2516
· www.promega.com
Part# TM274
Printed in USA.
Page 18
Revised 6/08
1. Press the “Menu” button to get to the Menu screen.
2. Scroll down to move the cursor to “Setup”. Press “Run/Stop” to
select “Setup”.
5314MA
1.
2.
Demo
3.
Setup
Run
====Menu====
3. Scroll up/down to move the cursor to the desired operational
mode. Press “Run/Stop” to select.
Refer to Table 1 for information on the operational mode required
for each Maxwell
®
16 Purification Kit.
5840MA
1.Research Mode
2.Forensic Mode
3.Cancel
===Setup===
4. Scroll down to move the cursor to the desired hardware mode.
If the instrument is configured with the Standard Elution
Volume hardware, select “SEV”. If the instrument is configured
with the Low Elution Volume hardware, select “LEV”.
Press “Run/Stop” to select.
6035MB
1.SEV Hardware
2.LEV Hardware
==Hardware Setup==
!
1. Make sure the instrument door is closed and turn on the Maxwell
®
16 Instrument. While the instrument is
homing itself during power up, the screen will display both the firmware version number and the
instrument’s current operational mode setting.
2. Verify that the current operational mode displayed matches the
operational mode required by both the installed hardware and the
purification kit to be used (see Table 1).
3. If the operational mode displayed does not match the operational
mode required by the purification kit, you will need to change the
instrument setting (see (ii) below, “
Changing the Maxwell
®
16
Instrument Operational Mode”
).
6033MA
Fnsc
Promega MAXWELL 16
Purification System
Version 4.0
LEV
Operational Mode Setting
Rsch = Research
Fnsc = Forensic
SEV = Standard Elution Volume
LEV = Low Elution Volume
tm274.0608.qxp 6/4/2008 9:52 AM Page 18