General Inspection and Maintenance
This section to be only completed by a qualified service technician as authorised by Prism
Medical UK . Use the provided service record history sheets to note findings during inspection.
A) Each Use - To be completed by User
All functions on hand control are operational.
B) Semi-Annually - To be completed at least every 6 months.
Should any of these items fail the inspection do not use the hoist. Contact Prism Medical UK or your
local qualified service technician for service.
Complete the inspection as noted in the “Each Use” section above.
Also check the following:
Battery charger functional.
Weigh scale functional (if applicable).
Castors are clean and free of debris.
All bolts and attachment points are secure (see next page for “Points of Attachment”)
Visual inspection of all weld points and cast components.
Check operation of the emergency stop button.
C) Annually - To be completed at least every 12 months.
Complete the inspection as noted in the “Semi-Annually” section above.
Also check the following:
Perform a working load test of one (1) lifting cycle with the maximum load. This is in accord-
ance to the periodic inspection requirements of ISO10535.
Check functional operation of the emergency lowering mechanism.
All bolts in connection with the carry bar must be inspected, greased and tightened. All defec-
tive or worn components must be replaced.
Connection bolts/bearings between arm and mast to be inspected. All defective or worn com-
ponents must be replaced.
Leg spreading function to be tested and all bolts to be inspected. All defective or worn com-
ponents must be replaced.
Check that main actuator is functional.