4 Adjusting the Condenser Settings |
4.3 Adjusting the Köhler Illumination
30
Primotech | Document 430055-9044-008 V.1.0; 03 / 2014
5
Lower the condenser until the edges of the luminous field diaphragm are in
focus.
6
Turn the condenser centering screws until the bright area is centered in the
field of view.
7
Open the luminous field diaphragm until the edge just disappears out of the
field of view.
8
Remove one of the eyepieces and look directly into the eyepiece tube.
This enables you to see the edges of the aperture diaphragm in focus.