7 Maintenance and Disposal
7.1 Routine Cleaning and Care
The maintenance to be carried out by the customer is limited to the following
activities:
¢
Cleaning the
external
optical surfaces of the following components
– Objectives and eyepieces
– Condenser
– Polarizers/analyzers and compensators
– All other glass surfaces
¢
Cleaning all other surfaces
INFO
Do not clean the internal optical surfaces. Touching or cleaning these surfaces
may lead to serious damage to the optical system.
Cleaning the Optical Surfaces
To clean the
external
optical surfaces of components:
Procedure
1
Switch the device off completely and pull the mains plug.
2
Remove the component from the microscope or move the stage so that the
component can be accessed.
For a list of components that can be removed and the corresponding
instructions, see Disassembling the Microscope [
47].
3
Gently rub the optical surfaces with an optical cleaning cloth dipped in an
optical cleaning solution (mixture of 85% petroleum ether and 15%
isopropanol).
– Rub the surfaces in a circular motion from the center outwards.
– To avoid scratches, do not use dry lens paper or a dry cloth.
– Make sure that no fluid enters the system.
4
Once the optical surfaces have dried, reinsert the component into the
microscope.