4 Adjusting the Condenser Settings |
4.2 Specifying the Condenser Position
28
Primotech | Document 430055-9044-008 V.1.0; 03 / 2014
Maximum Height
If desired, you can prevent the condenser from colliding with the bottom of the
sample by specifying the maximum height (Z position) of the condenser:
Procedure
1
Insert the 0.5 mm hex key in the condenser clamp screw.
2
Loosen the condenser clamp screw.
3
Look at the stage from above and carefully raise the condenser until the tip is
just below the sample.
4
Tighten the condenser clamp screw.
The condenser cannot move above this upper limit.