AFV-P Series High Performance Programmable AC Power Supply
99
6.3
HI-Range Voltage 60V
At the CALIBRATION page 1, users are allowed to enter into the page which calibrates
the HI-Range voltage 60V. The procedures of calibrating the HI-Range voltage 60V are
given as below:
1. Press the item
repeatedly to enter into the page
which calibrates the HI-Range voltage 60V (refer to Figure 6.9).
2. Connect the product with the voltage meter (refer to Figure 6.1)
3. Press the output & reset button on the front panel to enable the calibration of
the HI-Range voltage 60V (refer to Figure 6.10), and then the product will start to
output the voltage which is closed to 60V.
4. Use the rotary knob to adjust the product output until the measurement read-
ing of the output voltage shown on the touch screen is closed to the measure-
ment reading shown on the voltage meter.
5. Press the rotary knob to confirm and finish the calibration.
Figure 6.9 Enter into the page which calibrates the HI-Range voltage 60V
Figure 6.10 Enable the calibration of the HI-Range voltage 60V
NOTICE
Before calibrating the HI-Range voltage 60V, the load shall be temporally re-
moved from the product to avoid a potential electric shock.