27
WARNING
• The Nasal Cannula is intended for single patient use only. Use on more than
one patient may cause cross contamination.
• Use of accessories or replacement parts not listed in this User Manual may
cause adverse effects to basic safety or essential performance of the TOC.
•
Use humidifier bottle with TOC in the Continuous Flow mode only.
Use of
the humidifier bottle in the Pulse mode will cause the TOC to go into an
alarm condition.
• Prior to operation, ensure the inlet and exhaust vents on the TOC are clear. Any
blockage of the vents can inhibit performance.
• DO NOT place the TOC in a small closed space such as a closet.
• DO NOT cover the air vents, the TOC requires proper ventilation.
• ALWAYS keep at least 6 in. (15.24 cm) away from walls, furniture, and
especially curtains that could prevent adequate airflow to the TOC.
• In the event of an audible alarm or if you are experiencing any signs of
discomfort, connect to another oxygen source. DO NOT attempt to repair the
TOC. Contact your Provider and/or Healthcare Professional immediately.
• To ensure receiving the therapeutic amount of oxygen delivery according to your
medical condition, the TOC must be used only after one or more settings have
been individually determined or prescribed for you at your specific activity levels
by your Healthcare Provider.
• Use only parts recommended by the manufacturer to ensure proper function
and to avoid the risk of fire and burns.
• Use of this TOC at an altitude above 10,000 ft (3048 m) or outside a
temperature of 41°F to 104°F (5°C to 40°C) or a relative humidity above 95%
is expected to adversely affect the flowrate and the percentage of oxygen and
consequently the quality of therapy.
• ALWAYS confirm prescribed dose before use and monitor on a frequent basis.
• For proper oxygen flow, be sure that cannula is not kinked or obstructed before
or during use.
• Patients unable to communicate discomfort can require additional monitoring
and/or a distributed alarm system to alert caregiver of discomfort, medical
urgency or impending harm.
• The TOC contains magnetic, ferrous material that may affect the results of an MRI.
• Ensure TOC is properly engaged onto Cart before transporting.
• While transporting the TOC on the Cart avoid rough surfaces, holes and bumps.
• Use transportable Cart designed for TOC only. Do not use transportable Cart to
move other objects such as luggage.
• Do not modify the TOC or Cart in any way. Modifications could result in hazards
to the user.