9
Charging Battery with DC Power Cord
CAUTION
• Ensure the automobile power socket and the Auto Accessory Plug
connect properly.
• When powering the TOC in an automobile, ensure the vehicle’s engine is
running first, before connecting DC Power Cord into Auto 12 V Accessory Jack.
Operating the TOC without the engine running may drain the vehicle’s battery.
• Ensure that the Auto 12 V Accessory Jack is adequately fused for the TOC
power requirements (10 Amp). If the power socket cannot support a 10 Amp
load, the automobile’s fuse may blow and/or the socket may be damaged.
• It is not recommended to use a power inverter with the TOC.
DC Power Cord:
The DC Power Cord connects the TOC to an automobile’s
12 VDC outlet. The DC Power Cord will power the TOC and simultaneously
charge the Battery.
Connection:
Start the vehicle, connect the DC Power Cord to TOC (See
Connection of Power Cords to TOC for details) . Insert the Auto Accessory
Plug into automobile’s 12 V Accessory Jack.
Auto Accessory Plug
TOC Plug
connected to
TOC Receptacle
To replace fuse,
unscrew ribbed
washer, remove
fuse, replace with
same type and
rating fuse.
Auto 12 V
Accessory Jack