24
Operating Environment Ranges
Temperature: TOC: 41°F to 104°F (5°C to 40°C)
Humidity: 0% to 95%, non-condensing but not requiring a water
vapour partial pressure greater than 50 hPa
Atmospheric Pressure: 700 hPa to 1060 hPa
Altitude: 0 to 10,000 ft (0 to 3048 m)
Shipping and Storage Environmental Ranges
Temperature: -4˚F to 140˚F (-20˚C to 60˚C)
Humidity: 0% to 95%, non-condensing 95°F (35 °C) at a water vapour
pressure up to 50 hPa
Cannula
Pulse Mode: Maximum 7 ft (2.1 m) long standard adult single lumen
nasal cannula. Maximum oxygen flow ≥ 2 L/min
Continuous Mode: Maximum 30 ft (9.1 m) long standard adult single lumen
nasal cannula. Maximum oxygen flow ≥ 2 L/min
Expected Service Life of the device and accessories:
5 Years
Classifications
Mode of Operation: Continuous Duty
Type of Protections
against Shock: Class II
Degree of Protection
against Electrical
Shock:
Type BF
Degree of Protection
against Ingress of
Water:
IP22
VOLATILE ORGANIC COMPOUND (VOC) AND
PARTICULATE REQUIREMENTS
The oxygen delivered from the TOC meets the following requirements for particulate
levels, VOC levels, carbon monoxide levels, carbon dioxide levels and ozone levels.
EPA PM 2.5: Particulate Matter
ASTM D5466: VOC Levels
21 CFR 801.415: Ozone Levels
EPA NAAQS: Carbon Monoxide Levels
OSHA Permissible
Exposure Limits: Carbon Dioxide Levels
Standard Test Method for Determination of Volatile Organic Chemicals in
Atmospheres (Canister Sampling Methodology)