10
Connecting Nasal Cannula
to Oxygen Outlet Fitting
NOTE:
The amount of power provided from the vehicle’s DC electrical outlet
is limited, the power available to charge the Battery is determined by
the extra power available based on the operating demands of
the TOC.
At high flow settings, the extra power available to charge the Battery
may be limited and may increase the time to charge the Battery. At
the highest settings, there may not be any extra power available to
charge the Battery.
However, it is still a good practice to attach the TOC to a DC power
source when available.
START UP PROCEDURE
Checking for Proper Operation:
1. Connect the nasal cannula to the Oxygen Outlet Fitting of the TOC.
2. Turn on the TOC by selecting 2 L/min Continuous.
3. Verify that you hear an audible beep as the Control Knob is turned.
4. Verify the Battery Indicator LED’s show the status level of the Battery.
5. Verify the Continuous Mode LED’s are lit.
6. Gas should be flowing freely to the nasal cannula. You should be able
to hear or feel the flow of gas through the prongs of the nasal cannula.
Wave your hand in front of the prongs. If you do not feel the gas
flowing, check the cannula connections for leaks.