
11
Battery Indicator
Control Knob
Oxygen Outlet
Fitting
Continuous
Mode Settings
Pulse Mode
Settings
OFF
Alarm Light
Pulse Mode
LED’s
Continuous
Mode LED’s
Alarm Silence
Button
1. Choose a power option: AC Power Supply, DC Power Cord
or Battery.
• For AC Power Supply, see Charging Battery with AC Power Supply for
connection instructions.
• For DC Power Cord, see Charging Battery with DC Power Cord for
connection instructions.
• For Battery operation, verify battery status (level) prior to each use, charge
as necessary. See label on back of Battery for initial charging and installation
instructions. Ensure Battery is fully charged with AC Power Supply.
2. If running the TOC in PULSE mode:
• Connect the end of a standard adult single lumen oxygen nasal cannula with
a maximum length of 7 feet (2.1 m) to the TOC’s Oxygen Outlet Fitting.
CAUTION
• Inspect the TOC for visual damage before use, DO NOT USE
if damaged.
• Ensure Battery is fully charged prior to first use.
• To prevent damage to the TOC, DO NOT operate the TOC without the Air
Inlet Filter or while the Air Inlet Filter is wet.
• The Control Knob does not rotate 360°. Rotating the knob beyond the
stops will damage the TOC.
Starting at OFF position (12 o’clock):
• Rotate counter-clockwise for 1 to 5 Pulse Settings
• Rotate clockwise for 1/8 to 2 L/min Continuous Flow Settings
•
THERE IS NO FLOW BETWEEN SETTINGS.
• When the Battery is low (red light) replace the Battery with a charged
Battery or connect the TOC to AC or DC power source.
• Prior to operation, ensure the inlet and exhaust vents on the TOC are
clear. Any blockage of the vents can inhibit performance.
START UP PROCEDURE (continued)