CAUTION
• Do not use a humidifier bottle when pulse mode is selected.
• Irritation to the nasal passageways may occur with prolonged use
of oxygen.
3. If running the TOC in Continuous Mode:
• Connect the end of a standard adult single lumen oxygen nasal cannula with
a maximum length of 30 feet (9.1 m) to the TOC’s Oxygen Outlet Fitting.
• NOTE: If using humidifier bottle, Precision Medical, Inc. highly recommends
the PM500 Disposable Humidifier.
○ Using tubing adaptor P/N 508333, connect tubing from TOC Oxygen
Outlet Fitting to Humidifier Bottle Inlet Fitting.
○ Connect Cannula to Humidifier Bottle Outlet Fitting.
4. Place the cannula over your ears and position
the prongs in your nose as instructed by your
Equipment Provider or cannula manufacturer.
• NOTE: The proper placement and
positioning of the prongs of the nasal
cannula in the nose is critical to the amount
of oxygen delivered to the patient.
5. Turn ON the TOC by selecting the setting for your prescribed flow and mode
of operation.
• Turn the Control Knob left to select a “PULSE” setting and to the right to
select a “CONTINUOUS” setting.
OR
Turn Control Knob to
prescribed setting
6. Breathe normally through your nose.
• If running the TOC in PULSE mode, a measured pulse of oxygen will be
delivered each time the TOC senses an inhalation. To show the TOC is
triggering properly, the Pulse Mode LED’s will turn off briefly each time it
senses an inhalation.
• If running the TOC in CONTINUOUS mode, a measured continuous flow of
oxygen will be delivered. To show the TOC is in the CONTINUOUS mode,
the Continuous Mode LED’s will be
on steady.
7. To turn OFF the TOC, select the OFF position.