22
CLEANING, CARE and MAINTENANCE
CAUTION
• Prior to cleaning, ensure the TOC is turned off, unplug any external
power sources and remove Battery.
• DO NOT disassemble or attempt to repair. There are no user serviceable
parts inside. Contact your Provider and/or Healthcare Professional for
service.
• DO NOT spray or apply any cleaners directly onto the case.
• DO NOT allow any liquids to enter inside case.
• DO NOT use harsh and/or flammable chemicals to clean the TOC.
• DO NOT run the TOC until it is thoroughly dry.
• To prevent damage to the TOC, DO NOT operate the TOC without the
inlet filter or while the inlet filter is wet.
Perform cleaning as needed:
1. The patient should connect to an alternate oxygen source.
2. Select the
OFF
position on the TOC.
3. Unplug the external power source and remove battery before cleaning.
4. Clean exterior surfaces of the TOC with a cloth dampened with
mild detergent.
5. Wipe and allow TOC to air dry.
Note: When not in use, store the TOC in a clean dry area free from grease,
oil and other sources of contamination.
Inlet Filter:
1. Remove the inlet filter.
2. Wash filter with mild detergent. Rinse thoroughly with water and allow
to dry completely.
3. Once filter is dry, push the filter back into the case so only the small tab
is protruding.