19
Low Priority Alarms:
The following low priority alarm messages are accompanied by a
double beep
and a
solid yellow light.
Alarm Condition
Fault Code(s)
Possible Cause / Recommended Solution
Battery Low (15%)
FC1
Battery near depletion / Replace with charged battery
or switch to alternate power source.
Battery Charge
Temperature Low
FC5
1. Allow battery to warm up and charging will start
automatically after alarm is acknowledged.
2. Replace with another battery.
3. If alarm condition persists, contact your Healthcare
Provider. Switch to an alternate oxygen source.
Battery Charge
Temperature High
FC6
1. Allow battery to cool, charging will start automatically
after alarm is acknowledged.
2. Replace with another Battery.
3. I If alarm condition persists, contact your Healthcare
Provider. Switch to an alternate oxygen source.
Medium Priority Alarms:
The following medium priority alarm messages are accompanied by a
triple beep
repeated every 25 seconds and a
flashing yellow light
.
Alarm Condition
Fault Code(s)
Possible Cause / Recommended Solution
Oxygen Less than
83%
FC17
Oxygen Concentration is below 83% / Continue to use
TOC and contact your Healthcare Provider. Switch to
an alternate oxygen source.
High Temp Warning
FC7, FC8, FC9
TOC or parts of TOC are overheating / Power off TOC
and allow to cool. Ensure the TOC is well ventilated.
Return the TOC to your prescribed dose. If alarm
condition persists, contact your Healthcare Provider.
Switch to an alternate oxygen source.
Cannula Blocked
FC15
Cannula blocked or pinched / Verify the cannula is not
blocked or pinched. Replace cannula if damaged.
No Breath Detected
FC13
TOC is unable to detect patient’s breath with TOC in
Pulse mode / Verify cannula length is a maximum of
7 feet. Verify cannula connection is tight. Verify cannula
nasal prongs are properly placed into nostrils. Breathe
through the nose and not the mouth. Use “nasal
cannula” only; do not use cannula with mask. If alarm
condition continues, contact your Healthcare Provider.
Switch to an alternate oxygen source.
Battery Weak (7%)
FC2
Battery near depletion / Replace with charged battery
or switch to alternate power source.
Breath rate exceeding
35 breathes / minutes
FC14
1. Reduce level of activity to lower breath rate.
2. If alarm persists, change to alternate oxygen source
and contact your Equipment Provider.