Trident Model PD765 Universal Input Meter Instruction Manual
18
Thermocouple and RTD Connections
The following figures show examples for thermocouple and RTD con-
nections.
The RTD/TC selector switch must be set to the proper position for the
meter to accept the selected temperature input.
The input type is selected using the
Setup
menu.
Selected thermocouple input must correspond to thermocouple sensor
and wire type used.
Figure 8. Thermocouple Input Connections
Figure 9. Three-Wire RTD Input Connections
The meter accepts two, three, or four-wire RTDs. The three-wire RTD
connection has built-in lead wire compensation.