16 Maintenance and servicing
40
Regularly dismantle the weighing pan and the weighing pan holder and remove any dirt or dust from under
the weighing pan and on the balance housing with a soft brush or a soft, lint-free cloth, moistened with a
mild soap solution.
The weighing pan and the holder can be cleaned under running water. Take care that both parts are com-
pletely dry before they are re-installed on the balance.
The regular maintenance of the balance by your Service Representative will guarantee unlimited function and
reliability over many years and will extend the lifespan of the balance.
16.4 Error messages
The balance shows a description of the fault in the info-line.
16.4.1 Notes on correcting faults
The following table shows faults and their possible causes. If you cannot clear the fault using the table,
please contact the Customer Service..
WARNING
Never use solvents, acids, alkalis, paint thinners, scouring powders or other aggressive or corrosive
chemicals for cleaning, since these substances attack the surfaces of the balance housing and can cause
damage.
NOTE
If an error occurs without a description of the error in the info-line, the Customer Service must be called.
Fault
Possible cause
Weight dis-
play does not
light
• Balance not switched on
• Connection to power adaptor is interrupted
• Power supply has failed (interruption to current)
• The power adaptor is defective
“OL” is shown
in display
• The weight range has been exceeded (Observe information on the maximum weight
range)
“UL” is shown
in display
• The weight range is below the range of the balance (Scale pan or scale pan holder
missing)
The weight
display
fluctuates
continuously
• The draft is too strong at the balance location
• The balance support is vibrating or varying
• The scale pan is touching a foreign body
• The time chosen for FLOATINGDISPLAY is too short
• The material being weighed is absorbing moisture
• The material being weighed is being blown away or is evaporating
• Strong temperature variations in the material being weighed
Results of
weighing are
clearly incor-
rect
• The balance was not correctly tared
• The balance is not correctly levelled
• The calibration is no longer correct
• There are strong temperature variations
There is no
display or only
dashes
• The stability control (Balance functions) is set too sensitively
• The time selected for „FLOATINGDISPLAY“ is unsatisfactory
Configuration
menu cannot
be changed
• The password lock is activated in the configuration menu
The display
flashes contin-
uously during
calibration
• The balance location is not stable enough (Interrupt calibration with «
ON/OFF
» and re-
locate the balance in a better position)
• Use of an inprecise calibration weight (only applies to external calibration)