10 Working with the application menu
28
Operation:
Start the application (see Chapter 10.3 “Changing application / weighing mode”).
• Place the animal on the weighing pan.
• Press the «
REF
»-button until the desired function appears on the display.
• Release the «
REF
»-button.
10.11 Application “DENSITY”
See Chapter 8.5 “Activating the two main menus” and Chapter 8.6 “How the menu control operates” for the
activation of the Application menu.
You can carry out density determinations with the aid of the “DENSITY” program. For this, different modes
of density determinations can be selected.
•
Functions of the application on initialization:
•
Functions of the application on measurement:
Operation:
Start the application (see Chapter 10.3 “Changing application / weighing mode”).
• Press the «
REF
»-button until the desired function flashes on the display.
• Release the «
REF
»-button.
The program leads the operator through the individual operating steps.
10.11.1 Density determination “Solid on bottom mode”
A temperature-controlled reference-liquid (water) is poured into a bowl, placed on the scale pan and tared.
The solid is then placed into the liquid and weighed. Subsequently, the solid is suspended so that it is still
completely immersed in the water, but no longer touches the base. It is weighed again.
• SETUP APPLICATION
DENSITY
MODE
SOLID ON BOTTOM
MODE
SOLID IN AIR
MODE
LIQUID
MODE
SOLID POROUS
Solid body
Solid (under-floor)
Measure liquids
Solid porous bodies
INDEX
ON/
OFF
Index on/off
REFERENCE
8.000
Reference for index
TIMEBASE
0.0
Time base for repeat in seconds
REF. DENSITY
0.9988205
Density of the liquid used for the measure-
ment (set at works for water at 20°C)
TEMPERATUR
20.0 C
Temperature of the water used for the
measurement
NOTE
A Density Measurement Set (see Chapter 4 “Acessories”) is available as an optional extra. But you can
also determine the density of solids without this accessory.
«OK»:
- Accept current reference density
«CAL»:
- Calculate reference density of the measuring liquid
«T-H2o»: - Set reference density to water at nn.n °C
«20.0C»: - Set reference density to water at nn.n °C
«AIR», etc
- Invitation to measure the corresponding value
«<–>»:
-“Switch” from index to density
«SET»:
- Executes corresponding step
«STO»:
- Stores corresponding value (statistics)