11 Statistics
■
31
En
gl
is
h
Operation:
Start the application (see Chapter 10.3 “Changing application / weighing mode”).
«INF»:
- Change the display to the info display.
Displayed information:
- „Average value (MEAN)"
- „Standard deviation (STDE.)"
- „Relative standard deviation (STDE.-%)"
- „Maximum (MAX.)"
- „Minimum (MIN.)“
- „Total (TOT.)"
- „individual values"
- Exit info with «
MODE
»-button.
Display
Step
+
3,4780
9
Press «
MODE
» until
„STATISTIC“ appears.
statistic
+ 3,4780
9
Press «
REF
» shortly to store
the first stable measurement.
sto 0
res
inf
+ 3,4785
9
For „MANUAL“ recoding press «
REF
» shortly to store
a second measure-
ment.
For „
LOADCHANGE“ just place a new load.
For „
TIMEBASE“ wait until the time period set is over
sto 1
res
inf
+ 3,4775
9
Store a third
measurement
in the same manner as for the second
measure-
ment.
sto 2
res
inf
Retrieve the parameters:
+ 3,4775
9
Press «
REF
» until
„INF“ flashes
to s
how the info display with:
Average measurement
mean
3.4780 9
+ 3,4775
9
Standard deviation
STDe.
0.0005 9
+ 3,4775
9
Relative standard deviation
stde.-%
0.01 %
+ 3,4775
9
Maximum
max
3.4785 9
+ 3,4775
9
Minimum
min
3.4770 9
+ 3,4775
9
Minimum
TOT.
10.4340 9
+ 3,4775
9
Measurement 1
1= 3.4780 9
+
3,4775
9
Measurement 2, etc.
2= 3.4785 9
+ 3,4775
9
Exit info display with
«
MODE
»
sto 3
res
inf