Version 2.0
70
6_26931B ELS2100_2100 Ice Se rvice Manual_vsn2.0_maste r.doc
Programming Device Identity
Tools re quire d:
TERMINAL.exe or Hyperterminal.exe (supplied with
Windows)
ELSD Device ID.TRM program
Serial cable (supplied with instrument)
Parts re quire d:
N/A
1. Close down the control software
2. Connect the serial cable to the RS232 port on the
rear of the instrument
3. Start the
ELSD Device ID.TRM
program
(Figure 8) and configure the
Communications
in the
Settings
menu (figure 9)
4. Check the instrument successfully communicates
with the terminal program by selecting
Get Data
(figure 10).
5.
Now obtain the identity of the ELSD by
selecting
GET ID
.
The following value should be
obtained
001B000000000
X
B6
,where
X
is the
device ID (figure 11)
6.
To set the ELSD device ID to 1, select the
SET
ID 1
7.
Check the device ID has been set to the correct
value by selecting
GET ID.
The output should read
“001B000000000
1
B6”