Replacing the Nebuliser
When required:
If Nebuliser becomes blocked
Tools required:
Allen keys
Spanners
Parts required:
Nebuliser Assembly (# 0860-0401)
Ensure that the detector has cooled down before proceeding
Ensure that all gas and solvent connections are
disconnected prior to proceeding
1) Open the detector and remove the front
panel as described in
“Removing the front
panel”
(pg 15).
2) Remove the nebuliser from the
nebuliser mount using a spanner as
shown.