Removing And Cleaning The Evaporator Cartridge (Diffuser)
When required:
Loss or reduction in sensitivity. Noisy Baseline
Tools required:
Allen Keys
Parts required:
Evaporator Cartridge (# 0860-0425)
Ensure that the detector has cooled down before proceeding
1) Open the detector’s side panel
using the access key, as shown.
2) Remove the nebuliser chamber
as described in
“Replacing the
nebuliser chamber”
(pg 22).