Replacing The Vapour Sensors
When required:
Faulty Vapour sensor(s)
Tools required:
Access key
Small screwdriver
Parts required
Vapour Sensor (# 0860-0415)
The PL-ELS 2100/2100 Ice is fitted with a differential
vapour sensor system. The external sensor is located at the
rear of the instrument, the internal sensor is located on the
main control board
1) Open the detector’s side
panel using the access
key, as shown.
2)
Locate the external
vapour sensor on the rear
of the instrument.