Version 2.0
59
6_26931B ELS2100_2100 Ice Se rvice Manual_vsn2.0_maste r.doc
Replacing The PL-ELS 2100 Power Supply
When required:
Defective or Failed Power Supply
Tools required:
Screwdriver
Allen Keys
Parts required:
PL-ELS 2100 Power Supply (# 0860-1320)
Electronic boards and components are sensitive to electronic
discharge (ESD). In order to prevent damage always use ESD
protection whe n handling electronic boards and components.
1) Open the detector’s side
panel using the access key.
2) Unplug all connectors and
remove the two screws
indicated in the diagram
Carefully, slide the power
supply out from underneath
the guard
Insert the new Power supply
and reconnect the cables
Must be a shake-proof SEM
(SEM - Combined screw and
washer)