Version 2.0
60
6_26931B ELS2100_2100 Ice Se rvice Manual_vsn2.0_maste r.doc
Replacing The PL-ELS 2100 Ice Power Supply
When required:
Defective or Failed Power Supply
Tools required:
Screwdriver
Allen Keys
Parts required:
PL-ELS 2100 Ice Power Supply (# 0860-0640)
Electronic boards and components are sensitive to electronic
discharge (ESD). In order to prevent damage always use ESD
protection whe n handling electronic boards and components.
1) Open the detector’s side
panel using the access key.
2) Locate the power supply
on the main PCB.
Unplug the two connectors
highlighted