
Version 2.0
13
6_26931B ELS2100_2100 Ice Se rvice Manual_vsn2.0_maste r.doc
General Proble ms
Fault/Proble ms
Possible Cause(s)
Remedy
Baseline noise
The particle plume is not sufficiently
dried in the evaporator tube.
Increase the temperature of the
evaporator by 10ƒ intervals until the
noise is acceptable
Baseline noise
Poor nebulisation of solvent
Increase the gas flow rate
Decrease the nebulisation
temperature
Baseline noise
Pump pulsations, especially in
microbore applications where low flow
rates are used.
Use a pulse free pump
Increase the back pressure on
the pump by fitting a back pressure
regulator between the pump and the
injection valve.
Use a pulse dampener directly
after the pump in the system.
Baseline spikes
1. Particulate matter in the gas supply
2. Column shedding
3. Poor nebulisation
Filter the incoming gas, or
change the supply
Replace column or fit an inline
filter with a 0.2…m membrane filter
directly after the column.
Nebuliser temperature may be
too high and solvent may be boiling;
reduce nebuliser temperature
Low sensitivity and
baseline noise
Diffuser saturated with solvent
Stop the eluent flow and
increase the evaporator temperature
to 50ƒC above the current set
temperature. Increase the flow rate
to 2.8SLM and wait 15mins.
Large Baseline offset
1. Inefficient evaporation
2. High concentration of non-volatile
buffer or stabiliser
3. Contaminated diffuser
Increase the evaporator
temperature and/or gas flow.
Use a lower concentration of
stabiliser, unstabilised solvent or a
more volatile buffer
Perform ‘steam clean’.
(Section 4.2 Operation’s manual)
Peak tailing
Eluent particles lingering in the optical
chamber
Increase gas flow rate
Instrument Fails to zero
Offset too high or output unstable
Ensure the instrument is in
RUN mode
Refer to local distributor or
Polymer Laboratories
No power
1. Mains lead not connected
2. Fuse failure
Attach mains lead to socket and
inlet on rear of instrument
Replace fuse
No response
(completely flat
baseline)
1. Data acquisition leads not
connected
2. Light source inactive
3. Instrument in STANDBY mode
Ensure connectors to computer
or integrator are sound
Check LED
Select RUN mode
Temperature error as
soon as instrument
powered on
Temperature probe fault or
disconnected
Check RTD connections
Display not on, but
power connected, blue
glow ON.
Instrument in service mode
Rear panel switch-set to RUN