LED Check. Select each LED and observe Lobe to ensure correct LED displays. Combination of
LEDs will produce additive colours.
●
Sounder Check: Select Sounder and ensure Lobe Sounder functions.
●
Firmware version.
●
Bluetooth version.
●
Sensor check. Insert Sensor and move legs and ensure voltages change as appropriate.
System Check should be carried out once per year, or whenever a user reports a suspected functional
problem.
Tablet Restrictions
This menu allows the user to set the Tablet Lock option on or off. Enabling the Tablet Lock will restrict
this portable hand-
held device’s use to the RS App only. Attempting to access other portable hand
-
held device features and apps will return the user to the RS App.
Note
This functionality can only be used on iOS Systems.
Note
Certain basic functionality cannot be supressed. This includes the navigation button
swipe up menu and the system settings swipe down menu on iOS systems.
Installing Application from Distribution Service
Note
When deployed on an Apple device, the application will be installed and managed
by the manufacturer.
In some instances, the RS App may require reinstallation or updating. To do this, contact PMD device
solutions Ltd on [email protected]
Default Settings
The following table shows the default settings of the RR Monitor.
Table 12 Default Settings of the Respiratory Rate Mobile Medical Application
Option
Factory Default Setting
Configurable Setting/Status
Indicator
RR Default High Alarm Limit 24 breaths per minute
7 - 60
RR Default Low Alarm Limit
10 breaths per minute
6 - 59
RR High Alarm Limit
24 breaths per minute
7-60 (in use)
RR Low Alarm Limit
10 breaths per minute
6-59 (in use)
Averaging Window
15 minutes
2, 5 or 15 minutes
LED Brightness
350 LUX
Non-configurable
Status Indicator Volume
64 dB
Non-configurable
Alarm Signal Volume
64 dB
Non-configurable
Alarm Pause
Up to 50 seconds
Non-configurable