RespiraSense
™
Instructions For Use
RespiraSense
™
PDS-801-007 Revision 2
Page
40
of
77
1.
Connect to Lobe using procedure section.
2.
Select the “Lobe Settings” Button. The view will alter to present the Settings Interface.
3.
Alter LED Settings by pressing the LED Toggle Button to toggle them on or off.
4.
Verify that new settings are correct and p
ress the “OK” button to save the new settings to the
Lobe - the Sounder will beep and the Blue LED will flash.
5.
A summary will be displayed advising the user of the changes made to the indicator settings.
Press OK. The user will be returned to the Monitoring Screen.
6.
To instead discard altered settings, select the “Cancel” button. All changes will be discarded.
The user will be returned to the Monitoring Screen.
7.
Record alteration in patient records or in accordance with current hospital procedure.
8.
Confirm alteration of settings was successful:
●
System Sounder and LED icons are in the expected state.
●
Connection Sounder and LED indications will occur.
Step 10: Removal, Disposal, and Recharging
This Step presents procedures for dealing with the RS Device once patient monitoring has ended. The
RS Device must be removed when necessary and dealt with correctly after use. Remove both Lobe and
Sensor from patient. Do not leave sensor on patient if Lobe is not attached.
1.
The Lobe and Sensor must be removed when any of the following occur:
a.
Patient is discharged
b.
A period of 96 hours has elapsed since application
c.
The patient is undergoing a procedure which may damage the RS Device
d.
The patient is entering an area where environmental conditions may damage the RS
Device e.g. bathing or showering
e.
The RS Device becomes wet or soiled
f.
Any other situation which compromised patient safety or RS Device integrity.
2.
Depress the Lever on the cradle and push upwards in the space of the lobe to release the
Lobe from the Sensor
3.
Examine patient’s skin for sign of irritation or allergy.
4.
Gently Peel the sensor off patient’s skin with Lobe attached. If there is any difficulty in
removing the Sensor, first remove the Lobe, and then apply water to the fabric of the
sensor, or a hospital approved standard adhesive mover to the Sensor.
5.
Discard the Sensor in accordance with hospital, administrative, and/or local government
laws.
6.
Wipe down the Lobe in accordance with hospital, administration, and/or local government
laws.
7.
Return the Lobe to the Charging Station and ensure that the LED indicator shows a fully
charged status (Green) before re-use. Returning the Lobe to the Charging Station
terminates the current Lobe function and returns the Lobe to an unregistered state, with
no patient data with in it.
CAUTION
Dispose of Sensor in accordance with hospital administration and/or local
government laws. If in doubt, contact issuing organisation.