RespiraSense
™
Instructions For Use
RespiraSense
™
PDS-801-007 Revision 2
Page
53
of
77
●
Return to issuing organisation or department.
●
Seek replacement from Authorised Distributor.
Unable to scan Patient MRN barcode
●
Ensure camera lens is not obstructed or soiled.
●
Ensure there is sufficient light to scan.
●
Option: Patient MRN can be inputted manually
using keyboard.
●
Ensure camera is no more than 60 cm from
barcode.
●
Check patient barcode for damage or soiling.
●
Restart application and start scanning process
again.
●
Return to issuing organisation or department.
●
Seek replacement from Authorised Distributor.
Lobe ID not available on list of devices
using Bluetooth scanning option
●
Ensure Lobe is ready to be registered, a Green LED
should be flashing on a 3s cycle.
●
Option: Press Cancel and select Camera to scan
Lobe using camera interface.
●
Reset Lobe by returning it to the Charging Station.
●
Restart application and start scanning process
again.
●
Return to issuing organisation or department.
●
Seek replacement from Authorised Distributor.
Unable to enter Patient MRN using
keyboard
●
See "Tablet is unresponsive".
Default threshold limits are incorrect
●
Press “No” on confirmation screen to cancel
renaming. Default threshold limits are changed
using a password protected setting. Return RR
Monitor to authorised user to set default
thresholds. Restart renaming process.
●
If Yes was pressed on Confirmation screen see "
Threshold Limits are incorrect " in next sub-
section.
Incorrect patient MRN scanned during
patient registration
●
Return Lobe to Charging Station to reset. Restart
renaming procedure with correct MRN.
"Unable to Connect to Device." displayed
during patient registration
●
Reset Lobe by returning to Charging Station for at
least 5 seconds. Restart patient registration
process.
●
Return to issuing organisation or department.
●
Seek replacement from Authorised Distributor.
No Patient MRN barcode available
●
Select manual input option on the application.
Sounder does not beep at end of patient
registration process
●
Reset Lobe by returning to Charging Station for at
least 5 seconds. Restart patient registration
process.
●
Return to issuing organisation or department.
●
Seek replacement from Authorised Distributor.