RespiraSense
™
Instructions For Use
RespiraSense
™
PDS-801-007 Revision 2
Page
48
of
77
WARNING
The alarm thresholds should be scrutinised in light of the condition of the specific
patient and hospital procedures. Incorrect thresholds can lead to unnoticed
adverse events and failure to alert user to abnormal breathing rate. Only alter
threshold settings under medical supervision.
CAUTION
Verify that the upper and lower respiratory rate thresholds are in accordance with
current hospital, administrative, and/or local government law limits.
CAUTION
In cases of high or low respiratory rate readings the clinical status of the patient
should be assessed by a healthcare professional and/or supplemented with
additional physiological data.
CAUTION
All new applications of the RS Device will take the default thresholds. Verify that the
upper and lower RR thresholds are correct during renaming.
CAUTION
The respiratory rate thresholds for the alarm function of the Lobe have been pre-set
to standard industry limits. Ensure that these limits are correct prior to application
of the device.
Changing Password
Selecting the Change Password option from the Settings Menu allows the user to change the password
on the Settings Menu.
Password Complexity
Any password entered into the system must meet the following criteria:
●
Minimum of 8 Characters long
●
Minimum of 1 Uppercase Letter
●
Minimum of 1 Lower Case Letter
●
Minimum of 1 Number
●
Minimum of 1 Special Character
Set Result Screen Timeout
Selecting the Set Result Screen Timeout option from the Settings Menu allows the user to alter the
amount of time the Measurement Function Screen is displayed. This setting should be decided on with
the consultation of the end user and with consideration to patient confidentiality in the case the RS
Device is left unattended with a results screen displaying.
Note
This timeout is for the Results Screen only. All other screen timeouts are controlled
by the portable hand-held device settings.
System Check
Selecting the System Check option from the Settings Menu allows the user to perform system check
procedures. Tests consist of the following:
●
Accelerometer check. Move the Lobe and observe that accelerometer reading change.