CIP Training
Plockmatic LCT3500 Large Capacity Tray
Troubleshooting
Ultrasonic Sensor Calibration
B
Procedure - Calibration of Ultrasonic DSD
:
•
Select
Ultrasonic DSD Tray 3
to calibrate the
upper tray’s sensor;
•
Select
Ultrasonic DSD Tray 4
to calibrate the
lower tray’s sensor;
•
The selected tray opens;
•
Select
Optical DSD tray A
to calibrate the LSM
US sensor and open the LSM top cover;
•
Open the LCT door and fetch the Ultrasonic
DSD Calibration Strip [B];
•
Place the Calibration Strip as shown in the
pictures below. It shall be centered relative to
the US DSD Sensor;
•
Press
[Run];
•
Calibration complete. If calibration fails, old
settings will be kept.
US DSD Calibration Strip