![PKP DM08 Instruction Manual Download Page 32](http://html1.mh-extra.com/html/pkp/dm08/dm08_instruction-manual_1561318032.webp)
Instruction Manual
DM08
Service password:
The service password is not required for para-
meter settings which are relevant for operation.
The service password is reserved for service
technicians and not provided to customers.
8.1.3 Totalizer
The TOTALIZER functional class includes the
following functions:
For changing the current settings and paramet-
ers the customer password is needed.
Otherwise, the settings only can be watched but
not be changed.
The ESC key is used for discarding changes.
Totalizer unit:
For selecting the desired function please use
the key
▲
or
▼
.
After choosing the function
Totalizer unit
and pressing ENTER key, the current forward
and reverse totalizer unit will be displayed.
Totalizer unit
[ m3 ]
Example:
How to change the parameter please refer to
Section "Selection window / make a selection".
The following units are available:
Volume units
m3
Cubic metre
l
Litre
USG Gallon (US)
UKG Gallon (brit.)
ft3
Cubic feet
Mass units*
kg
Kilogram
t
Tonne
* When selecting a mass unit the liquid density is
certainly required ("Flow", p. 33).
The changes will be confirmed by pressing the
ENTER key and the new unit will be active as
well for the forward totalizers and the reverse
totalizers.
Notice
When the unit is changed, the totalizers
will be reset to 0.00 automatically!
Totalizer reset:
The display electronics of the DM08 has three
independent totalizers.
Totalizer 1 and Totalizer 2 for the forward flow
and a totalizer for reverse flow.
Technical changes reserved
-
31
-